J Korean Soc Magn Reson Med.  2009 Dec;13(2):161-170.

Optimizations of 3D MRI Techniques in Brain by Evaluating SENSE Factors

Affiliations
  • 1Department of Radiology, East-West Neo Medical Center, Kyung Hee University, Seoul, Korea. ghjahng@gmail.com
  • 2Department of Radiology, School of Medicine, Kyung Hee University, Korea.

Abstract

PURPOSE
A parallel imaging method provides us to improve temporal resolution to obtain three-dimensional (3D) MR images. The objective of this study was to optimize three 3D MRI techniques by adjusting 2D SESNE factors of the parallel imaging method in phantom and human brain.
MATERIALS AND METHODS
With a 3 Tesla MRI system and an 8-channel phase-array sensitivity-encoding (SENSE) coil, three 3D MRI techniques of 3D T1-weighted imaging (3D T1WI), 3D T2-weighted imaging (3D T2WI) and 3D fluid attenuated inversion recovery (3D FLAIR) imaging were optimized with adjusting SESNE factors in a water phantom and three human brains. The 2D SENSE factor was applied on the phase-encoding and the slice-encoding directions. Signal-to-noise ratio(SNR), percent signal reduction rate(%R), and contrast-to-noise ratio(CNR) were calculated by using signal intensities obtained in specific regions-of-interest (ROI).
RESULTS
In the phantom study, SENSE factor=3 was provided in 0.2% reduction of signals against without using SENSE with imaging within 5 minutes for 3D T1WI. SENSE factor=2 was provided in 0.98% signal reduction against without using SENSE with imaging within 5 minutes for 3D T2WI. SENSE factor=4 was provided in 0.2% signal reduction against without using SENSE with imaging around 6 minutes for 3D FLAIR. In the human brain study, SNR and CNR were higher with SENSE factors=3 than 4 for all three imaging techniques.
CONCLUSION
This study was performed to optimize 2D SENSE factors in the three 3D MRI techniques that can be scanned in clinical time limitations with minimizing SNR reductions. Without compromising SNR and CNR, the optimum 2D SENSE factors were 3 and 4, yielding the scan time of about 5 to 6 minutes. Further studies are necessary to optimize 3D MRI techniques in other areas in human body.

Keyword

3D MRI; 2D SENSE; 3D T1-weighted; 3D T2-weighted; 3D FLAIR

MeSH Terms

Brain
Human Body
Humans
Water
Water