Lab Med Online.  2023 Oct;13(4):308-317. 10.47429/lmo.2023.13.4.308.

Evaluation of Two Safety Syringes for Arterial Blood Gas Analysis

Affiliations
  • 1Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

Abstract

Background
Arterial blood gas (ABG) analysis is a crucial test providing vital information for patients who are critically ill. Accurate results are essential; however, pre-implementation evaluation of ABG syringes is uncommon. We evaluated the performance of the Greenmedi heparin syringe (GRS; SANG-A FRONTEC, Korea) and the Portex arterial blood sampling kit (POS; Smiths Medical, USA), which are widely used in clinical laboratories.
Methods
Samples from Asan Medical Center were assessed for clotting rate and hemolysis rate. Gas permeability, sample volume impact, precision, and concordance were evaluated using quality control materials. Gas permeability was assessed by measuring pO 2, pCO 2, and pH after storing the materials for different time periods. The influence of sample volume was tested at various quantities. Precision was evaluated through repeated tests, and concordance between the two syringes’ results was analyzed.
Results
GRS had a clotting rate of 1.07% (51/4,769), exceeding the acceptable range, whereas POS had no clotted samples (0/387). Hemolysis rates were 14.9% (14/94) for GRS and 10.0% (8/80) for POS, both higher than the acceptable rate. Both syringes had a storage time of less than 10 minutes and required a minimum sample volume above 0.6 mL. Precision was noteworthy, but there was a relatively weak concordance between pO 2 values obtained with the two syringes.
Conclusions
Both syringes demonstrated notable precision but had high hemolysis rates. Results showed low concordance and were influenced by storage time and sample volume. Caution should be exercised when using these syringes, and it is recommended to assess their quality before implementation in each laboratory.

Keyword

Blood gas analysis; Syringes; Evaluation

Figure

  • Fig. 1 Study flow diagram for arterial blood gas analysis of GRS and POS samples. (A) Clotting rate and (B) hemolysis rate. Abbreviations: GRS, Greenmedi heparin syringe; POS, Portex arterial blood sampling kit; ABGA, arterial blood gas analysis; ICU, intensive care unit; pHb, plasma hemoglobin.

  • Fig. 2 Passing–Bablok regression and Bland-Altman analysis of 120 QC materials. The pO2, pCO2, and pH results obtained using the Portex arterial blood sampling kit (POS) compared with those obtained using the Greenmedi heparin syringe (GRS) were plotted with the passing-Bablok regression and identity lines (left). The equations of the regression lines are shown (left) and the Bland-Altman plots are displayed on the right side. Abbreviations: pO2, partial pressure of oxygen; pCO2, partial pressure of carbon dioxide.


Reference

1. Lippi G, Panteghini M, Bernardini S, Bonfanti L, Carraro P, Casagranda I, et al. 2018; Laboratory testing in the emergency department: an Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC) and Academy of Emergency Medicine and Care (AcEMC) consensus report. Clin Chem Lab Med. 56:1655–9. DOI: 10.1515/cclm-2017-0077. PMID: 28361782.
2. Shapiro BA, Peruzzi WT, editors. 1994. Clinical Application of Blood Gases. 5th ed. Chicago, IL: Mosby.
3. Clinical and Laboratory Standards Institute. 2009. Blood gas and pH analysis and related measurements; Approved guideline-second edition. CLSI document C46-A2. Clinical and Laboratory Standards Institute;Wayne, PA:
4. Küme T, Sişman AR, Solak A, Tuğlu B, Cinkooğlu B, Coker C. 2012; The effects of different syringe volume, needle size and sample volume on blood gas analysis in syringes washed with heparin. Biochem Med (Zagreb). 22:189–201. DOI: 10.11613/BM.2012.022. PMID: 22838185.
5. Kim HS, Lee KC, Jung WS, Chang YJ. 2012; Effects of different sample volumes on arterial blood gas and electrolytes testing. Anesth Pain Med. 7:121–4.
6. Cho NS, KIM CH. 2000; The study on changes in arterial blood gas finding with time interval. J Korean Soc Emerg Med. 11:27–34.
7. Hess CE, Nichols AB, Hunt WB, Suratt PM. 1979; Pseudohypoxemia secondary to leukemia and thrombocytosis. N Engl J Med. 301:361–3. DOI: 10.1056/NEJM197908163010706. PMID: 460325.
8. Aminabhavi TM, Aithal US. 1991; Molecular transport of oxygen and nitrogen through polymer films. Journal of Macromolecular Science, Part C. 31:117–63. DOI: 10.1080/15321799108021922.
9. Schmidt C, Müller-Plathe O. 1992; Stability of pO2, pCO2 and pH in heparinized whole blood samples: influence of storage temperature with regard to leukocyte count and syringe material. Eur J Clin Chem Clin Biochem. 30:767–73. DOI: 10.1515/cclm.1992.30.11.767. PMID: 1489849.
10. Pretto JJ, Rochford PD. 1994; Effects of sample storage time, temperature and syringe type on blood gas tensions in samples with high oxygen partial pressures. Thorax. 49:610–2. DOI: 10.1136/thx.49.6.610. PMID: 8016801. PMCID: PMC474964.
11. New PF, Baker E. 1963; Technique of arterial puncture: a new needle-cannula for arteriography. J Neurosurg. 20:390–6. DOI: 10.3171/jns.1963.20.5.0390. PMID: 14186055.
12. Clinical and Laboratory Standards Institute. 2004. Procedures for the collection of arterial blood specimens; Approved standard-Fourth edition. CLSI document GP43-A4. Clinical and Laboratory Standards Institute;Wayne, PA:
13. Ministry of Food and Drug Safety. Guidelines for certification of blood collection sets and preparation of technical documents. https://www.mfds.go.kr/brd/m_1060/view.do?seq=14719&srchFr=&srchTo=&srchWord=%EC%B1%84%ED%98%88&srchTp=1&itm_seq_1=0&itm_seq_2=0&multi_itm_seq=0&company_cd=&company_nm=&Data_stts_gubun=C9999&page=1. Update on Sep 2020.
14. Chang J, Kim S, Yoo SJ, Park EJ, Um TH, Cho CR. 2020; Preanalytical errors in the central laboratory of a university hospital based on the analysis of year-round data. Clin Lab. 66:1783–91. DOI: 10.7754/Clin.Lab.2020.200110. PMID: 32902239.
15. Jang H, Cho Y, Rim JH, Kang H, Lee SG, Lim JB. 2021; Measurement of plasma free hemoglobin using hemolysis index and bilirubin interference. Lab Med Qual Assur. 43:208–13. DOI: 10.15263/jlmqa.2021.43.4.208.
16. Westgard QC. Desirable biological variation database specifications. https://www.westgard.com/biodatabase1.htm. Last accessed on Jan 2023.
17. European Federation of Clinical Chemistry and Laboratory Medicine. EFLM Biological Variation Database. https://biologicalvariation.eu/. Last accessed on Jan 2023.
18. Kofstad J. 1993; Quality specifications for pH and blood gases. Upsala Journal of Medical Sciences. 98:381–5. DOI: 10.3109/03009739309179336. PMID: 7974869.
19. Clinical and Laboratory Standards Institute. 2018. Measurement procedure comparison and bias estimation using patient samples. 3rd ed. CLSI guideline EP09c. Clinical and Laboratory Standards Institute;Wayne, PA:
20. Calaf N, Giner J, Codina E, Feixas T, González M, Casan P. 2004; Comparison of arterial blood sample kits. Arch Bronconeumol. 40:378–80. DOI: 10.1016/S0300-2896(04)75548-9. PMID: 15274868.
21. Bosch M, Bauzá F, Togores B, Agustí AG. 1997; Comparative evaluation of 3 types of syringes routinely used for arterial gas analysis. Arch Bronconeumol. 33:331–4. DOI: 10.1016/S0300-2896(15)30580-9. PMID: 9410433.
22. Park JM, Park WB, Kim JJ, Min KJ, Choi WS, Jang JH, et al. 2017; Comparison of the accuracy of syringes washed with heparin, manually and heparin coated product for Arterial Blood Gas Analysis on electrolytes and hemoglobin. Journal of The Korean Society of Emergency Medicine. 28:431–40.
23. Picandet V, Jeanneret S, Lavoie JP. 2007; Effects of syringe type and storage temperature on results of blood gas analysis in arterial blood of horses. J Vet Intern Med. 21:476–81. DOI: 10.1111/j.1939-1676.2007.tb02993.x. PMID: 17552454.
24. Knowles TP, Mullin RA, Hunter JA, Douce FH. 2006; Effects of syringe material, sample storage time, and temperature on blood gases and oxygen saturation in arterialized human blood samples. Respir care. 51:732–6.
25. Gershengorn HB, Garland A, Kramer A, Scales DC, Rubenfeld G, Wun-sch H. 2014; Variation of arterial and central venous catheter use in United States intensive care units. Anesthesiology. 120:650–64. DOI: 10.1097/ALN.0000000000000008. PMID: 24424071. PMCID: PMC4145875.
26. Cho GL, Choi JS. 2010; Knowledge of and compliance with standard precautions by nurses in intensive care unit. J Korean Acad Fundam Nurs. 17:73.
27. Yu JW, Yang NY. 2017; Characteristics and knowledge of needlestick injuries and compliance with standard precautions in healthcare workers. Journal of Korean Academic Society of Home Health Care Nursing. 24:275–83.
28. Lima-Oliveira G, Lippi , Salvagno GL, Montagnana M, Picheth G, Guidi GC. 2012; Different manufacturers of syringes: a new source of variability in blood gas, acid-base balance and related laboratory test? Clin biochem. 45:683–7. DOI: 10.1016/j.clinbiochem.2012.03.007. PMID: 22440459.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr