Ultrasonography.  2020 Jan;39(1):85-93. 10.14366/usg.19033.

Comparison of MicroFlow Imaging with color and power Doppler imaging for detecting and characterizing blood flow signals in hepatocellular carcinoma

Affiliations
  • 1Department of Radiology, Seoul National University Hospital, Seoul, Korea
  • 2Department of Radiology, Seoul National University College of Medicine, Seoul, Korea
  • 3Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Korea
  • 4Department of Radiology, Seoul Metropolitan Government-Seoul National University Boramae Medical Center, Seoul, Korea

Abstract

Purpose
The purpose of this study was to compare the sensitivity of MicroFlow Imaging (MFI) with that of color and power Doppler imaging (CDI and PDI, respectively) in detecting the vascularity of hepatocellular carcinomas (HCCs).
Methods
This prospective study enrolled 51 patients diagnosed with HCC between August 2018 and December 2018. CDI, PDI, MFI, and contrast-enhanced ultrasound (CEUS) were performed. Two radiologists evaluated the presence and pattern of tumoral vascularity on CDI, PDI, and MFI. Vascular presence was graded on a 5-point scale (0, absent; 4, >50% of the tumor). The vascular pattern was chosen from following categories: basket, vessels in tumor, spot, detouring, mixed, or others. Two additional radiologists assessed CEUS images for the presence and pattern of tumoral vascularity, which served as the reference standard. If the tumoral vascular pattern on each examination matched that of the CEUS images, the Wilcoxon test and McNemar test, respectively, were used to compare the sensitivity for detecting tumoral vascularity between MFI and CDI, and between MFI and PDI. Logistic regression analysis was performed to identify factors associated with MFI detectability of tumoral vascularity.
Results
CEUS demonstrated tumoral vascularity in 98.0% (50 of 51) of patients. MFI (58.0%, 29 of 50) demonstrated a higher sensitivity than CDI (14.0%, 7 of 50) or PDI (14.0%, 7 of 50) (P<0.001 for both) in detecting tumoral vascularity, provided that the vascular pattern was correctly depicted. Only tumor depth was associated with the MFI detectability of tumoral vascularity.
Conclusion
The sensitivity of MFI was higher than that of CDI or PDI in detecting the vascularity of HCCs when the vascular pattern was considered. MFI better detected the vascularity of shallow tumors.

Keyword

Ultrasonography; Ultrasonography, Doppler, color; Carcinoma, hepatocellular