J Korean Neurosurg Soc.  2009 May;45(5):284-288. 10.3340/jkns.2009.45.5.284.

Assessment of the Intracranial Stents Patency and Re-Stenosis by 16-Slice CT Angiography with Optimized Sharp Kernel : Preliminary Study

Affiliations
  • 1Department of Diagnostic Radiology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Busan, Korea.
  • 2Department of Diagnostic Radiology, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Korea. drcello@pusan.ac.kr
  • 3Department of Neurosurgery, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Korea.
  • 4Department of Neurology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Busan, Korea.

Abstract


OBJECTIVE
Our retrospective study aimed to determine whether 16-slice computerized tomography (CT) angiography optimized sharp kernel is suitable for the evaluation of visibility, luminal patency and re-stenosis of intracranial stents in comparison with conventional angiography.
METHODS
Fifteen patients with symptomatic intracranial stenotic lesions underwent balloon expandable stent deployment of these lesions (10 middle cerebral arteries, 2 intracranial vertebral arteries, and 3 intracranial internal carotid arteries). CT angiography follow-up ranged from 6 to 15 months (mean follow-up, 8 months) after implantation of intracranial stents and conventional angiography was confirmed within 2 days. Curved multiplanar reformations with maximal intensity projection (MIP) with optimal window settings for assessment of lumen of intracranial stents were evaluated for visible lumen diameter, stent patency (contrast distal to the stent as an indirect sign), and re-stenosis by two experienced radiologists who blinded to the reports from the conventional angiography.
RESULTS
All of stents deployed into symptomatic stenotic lesions. All stents were classified as patent and no re-stenosis, which was correlated with results of conventional angiography. Parts of the stent lumen could be visualized in all cases. On average, 57% of the stent lumen diameter was visible using optimized sharp kernel. Significant improvement of lumen visualization (22%, p<0.01) was observed using the optimized sharp kernel compared with the standard sharp kernel. Inter-observer agreements on the measurement of lumen diameter and density were judged as good, respectively (p<0.05).
CONCLUSION
Sixteen-slice CT using the optimized sharp kernel may provide a useful information for evaluation of lumen diameter patency, and re-stenosis of intracranial stents.

Keyword

Intracranial atherosclerosis; Stents; Multislice CT; CT angiography; Cerebral angiography

MeSH Terms

Angiography
Cerebral Angiography
Follow-Up Studies
Humans
Intracranial Arteriosclerosis
Middle Cerebral Artery
Phenobarbital
Retrospective Studies
Stents
Vertebral Artery
Phenobarbital
Full Text Links
  • JKNS
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr