J Lab Med Qual Assur.  2016 Dec;38(4):234-242. 10.15263/jlmqa.2016.38.4.234.

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Affiliations
  • 1Department of Laboratory Medicine, Hanil General Hospital, Seoul, Korea. jks3393@hanmail.net
  • 2Division of Endocrinology and Metabolism, Hanil General Hospital, Seoul, Korea.

Abstract

BACKGROUND
We evaluated the accuracy of the B. Braun Omnitest 5 blood glucose monitoring system (BGMS; Infopia Co. Ltd., Korea), which was recently developed for self-monitoring of blood glucose.
METHODS
Precision was assessed according to Clinical and Laboratory Standards Institute guideline EP5-A3 with control materials containing low, normal, and high levels of glucose. Linearity was evaluated over the range of 52.5-548.0 mg/dL prepared from patient samples. For system accuracy, 100 capillary blood samples measured by the B. Braun Omnitest 5 BGMS were compared to plasma-equivalent blood glucose values of the fingertip blood samples measured by the YSI 2300 STAT PLUS glucose analyser (YSI Life Sciences, USA). Accuracy was evaluated according to the International Organization for Standardization (ISO) 15197: 2013 criteria.
RESULTS
The range for the total coefficient of variation (%) was 1.5%-4.9% for three lots of strips. Both within-run precision and within-laboratory precision fulfilled the manufacturer's claim. ISO 15197: 2013 states that more than 95% of blood glucose measurement values must be within ±15 mg/dL for ranges below 100 mg/dL and ±15% for ranges above 100 mg/dL. Each of B. Braun Omnitest 5 lots satisfied ISO 15197: 2013, as 98.5% (197/200), 97.0% (194/200), and 99.5% (199/200) of values were within range. In consensus error grid analysis, respectively, 99.5%, 99.0%, and 100.0% of measurement values of each lot were within zone A, indicating that an average of over 99.0% of values were within zone A.
CONCLUSIONS
B. Braun Omnitest 5 provided reliable results and satisfied the ISO 15197: 2013 accuracy criteria. This test is an appropriate BGMS for the self-monitoring of blood glucose.

Keyword

Blood glucose monitoring system; Self-monitoring of blood glucose; Accuracy; Evaluation

MeSH Terms

Biological Science Disciplines
Blood Glucose*
Capillaries
Consensus
Glucose
Humans
Blood Glucose
Glucose
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr