J Korean Orthop Assoc.  1998 Aug;33(4):1134-1139.

Changes of Fixation Strength by Rod - Contouring of Compact Cotrel - Dubousset Instrumentation

Abstract

One of various decompression methods in treatment of spinal stenosis is the indirect instrumental decompression. Theoretically, the distraction of the disc space can widen the intervertebral foramen of the stenotic segment and even increase the canal diameter by distracting the posterior annulus as well as reduce the extent of decompressive laminectomy site. The indirect instrumental decompression, however, was not guaranteed to maintain the restored discal height because of the loss of fixation strength between rod and screw, viscoelasticity of vertebra itself, bone density, type of screw and rod, and operative technique. As well the magnitude of the stresses on the instrumentation particularly at the rod-screw interface may depend on rod-contouring in order to make mormal sagittal curvature of the lumbar spines. Therefore, the aim of this experimental study was to evaluate the effect of different rod-contour on the axial sliding strength in Compact Cotrel-Dubousset (CCD) instrumentation. Axial sliding strength was tested by Universal Test Machine (Instron). Test was performed for 3 groups of different rodcontouring on the biomechanical axial strength: straight rod (no contour), 10 and 20 contouring rod. The length of contact surface between rod and screw was measured with Fuji pressure sensitive film. The study was performed using 6.5 mm open body screws and 7 mm rods of CCD instrumentation. Axial sliding strength of straight rod was 2518.6N, 1871.8N in 10 and 1528.8N in 20 contouring rod. The length of contact surface between rod and screw significantly decreased according to degree of rod contouring; 9.88mm in straight rod, 9.08mm in 10 and 8.57mm in 20 contouring rod. There was a statistically significant linear correlation (R=0.96) between failure load and length of contact surface. Therefore, this study has shown that excessive contour of the rod in order to make normal sagittal curvature of the lumbar spine using CCD instrumentation cannot provide sufficient axial sliding strength. That may be a cause of loss of restored disc space height after surgery.

Keyword

CCD; Axial strength; Length; Rod contouring

MeSH Terms

Bone Density
Decompression
Laminectomy
Spinal Stenosis
Spine
Full Text Links
  • JKOA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr