Korean J Urol.  1999 Jul;40(7):905-908.

The Usefulness of RigiScan(R) as Screening Test for Duplex Ultrasonography after Intracorporeal Injection

Affiliations
  • 1Department of Urology, School of Medicine, Soonchunhyang University, Seoul, Korea.
  • 2Department of Radiology, School of Medicine, Soonchunhyang University, Seoul, Korea.

Abstract

PURPOSE: To investigate whether the result of RigiScan after intracorporeal injection could predict the result of duplex ultrasonography, and determine the necessity of duplex ultrasonography.
MATERIALS AND METHODS
We performed both RigiScan and duplex ultrasonography after intracorporeal injection in 18 men with erectile dysfunction. All of the patients were evaluated by history taking, physical examination, laboratory test, hormonal tests, and neurologic examination before RigiScan and duplex ultrasonography. We measured maximal arterial diameter, peak systolic velocity, end-diastolic velocity using a 7 MHz. color Doppler unit and these results of duplex ultrasonography were compared with penile rigidity and tumescence measured by RigiScan.
RESULTS
The positive and negative predictive value of RigiScan were 81.8% and 85.7%, respectively. The result of RigiScan was comparable with that of duplex ultrasonography and the accuracy of RigiScan was resonable.
CONCLUSIONS
According to this results, duplex ultrasonography was needed if the response of RigiScan is abnormal after intracorporeal injection. However, duplex ultrasonography can be ommitted if the response of RigiScan is normal after intracorporeal injection.

Keyword

RigiScan(r); Duplex ultrasonography; Erectile dysfunction

MeSH Terms

Erectile Dysfunction
Humans
Male
Mass Screening*
Neurologic Examination
Physical Examination
Ultrasonography*
Full Text Links