J Clin Neurol.  2020 Apr;16(2):314-320. 10.3988/jcn.2020.16.2.314.

Automatic Derivation of Continuous Positive Airway Pressure Settings: Comparison with In-Laboratory Titration

Affiliations
  • 1Department of Neurology, Catholic Kwandong University College of Medicine, International St. Mary's Hospital, Incheon, Korea.
  • 2Sleep Disorders Center, Department of Neurology, Soonchunhyang University College of Medicine, Cheonan Hospital, Cheonan, Korea.
  • 3Department of Preventive Medicine, Soonchunhyang University College of Medicine, Cheonan, Korea.

Abstract

Background and Purpose
This study was designed to investigate differences in the final recommended pressure setting between that derived from an autotitrating continuous positive airway pressure (APAP) device and manual in-laboratory continuous positive airway pressure (CPAP) titration, as well as the factors that influence pressure differences in patients with obstructive sleep apnea (OSA).
Methods
This study enrolled 50 patients with OSA. All patients underwent both APAP titration and manual CPAP titration. We obtained the average device pressure ≤90% of the time (APAP90) from the downloaded manual for the APAP machine and the optimal pressure obtained by manual CPAP titration (CPAPmanual). We placed the subjects into three groups based on the pressure difference (ΔP) obtained by subtracting CPAPmanual from APAP90: 1) Prequal (ΔP=0), 2) CPAPmanual+ (ΔP ≤−1), and 3) APAP90+ (ΔP ≥1). Regression analysis was conducted to identify predictive factors associated with ΔP.
Results
The values of APAP90 and CPAPmanual were 9.50±3.03 cmH2O and 9.48±2.71 cmH2O (mean±SD), respectively (p=0.95). The Prequal, CPAPmanual+, and APAP90+ groups comprised 9 (18%), 23 (46%), and 18 (36%) subjects, respectively. Regression analyses revealed that male sex [β=3.539, 95% confidence interval (CI)=0.040–7.039], body mass index (BMI) (β=0.186, 95% CI=0.020–0.352), and average usage per day (β=0.768, 95% CI=0.077–1.459) were associated with ΔP.
Conclusions
While the mean pressure in the overall cohort did not differ significantly between APAP90 and CPAPmanual, there was a discordance majority showing different single pressures obtained when applying the two titration methods. Being Male, having an increased BMI, and having an increased average usage per day of APAP were significantly correlated with increased ΔP in this study.

Keyword

obstructive sleep apnea; polysomnography; continuous positive airway pressure