Asian Spine J.  2017 Oct;11(5):733-738. 10.4184/asj.2017.11.5.733.

The “Skipped Segment Screw” Construct: An Alternative to Conventional Lateral Mass Fixation–Biomechanical Analysis in a Porcine Cervical Spine Model

Affiliations
  • 1Department of Spinal Disorders, Christian Medical College, Ida Scudder Road, Vellore, India. kedarorth@gmail.com
  • 2Division of Photonics, Vellore Institute of Technology, Near Katpadi Road, Vellore, India.

Abstract

STUDY DESIGN: Cadaveric biomechanical study. PURPOSE: We compared the "skipped segment screw" (SSS) construct with the conventional "all segment screw" (ASS) construct for cervical spine fixation in six degrees of freedom in terms of the range of motion (ROM). OVERVIEW OF LITERATURE: Currently, no clear guidelines are available in the literature for the configuration of lateral mass (LM) screwrod fixation for cervical spine stabilization. Most surgeons tend to insert screws bilaterally at all segments from C3 to C6 with the assumption that implants at every level will provide maximum stability.
METHODS
Six porcine cervical spine specimens were harvested from fresh 6-9-month-old pigs. Each specimen was sequentially tested in the following order: intact uninstrumented (UIS), SSS (LM screws in C3, C5, and C7 bilaterally), and ASS (LM screws in C3-C7 bilaterally). Biomechanical testing was performed with a force of 2 Nm in six degrees of freedom and 3D motion tracking was performed.
RESULTS
The two-tailed paired t-test was used for statistical analysis. There was a significant decrease in ROM in instrumented specimens compared with that in UIS specimens in all six degrees of motion (p<0.05), whereas there was no significant difference in ROM between the different types of constructs (SSS and ASS).
CONCLUSIONS
Because both configurations provide comparable stability under physiological loading, we provide a biomechanical basis for the use of SSS configuration owing to its potential clinical advantages, such as relatively less bulk of implants within a small operative field, relative ease of manipulating the rod into position, shorter surgical time, less blood loss, lower risk of screw-related complications, less implant-related costs, and most importantly, no compromise in the required stability needed until fusion.

Keyword

Cadaver; Cervical vertebrae; Laminectomy; Swine; Biomechanics

MeSH Terms

Cadaver
Cervical Vertebrae
Equidae
Female
Freedom
Laminectomy
Operative Time
Range of Motion, Articular
Spine*
Surgeons
Swine