Radiat Oncol J.  2011 Sep;29(3):199-205. 10.3857/roj.2011.29.3.199.

A comparison of preplan MRI and preplan CT-based prostate volume with intraoperative ultrasound-based prostate volume in real-time permanent brachytherapy

Affiliations
  • 1Department of Radiation Oncology, CHA Bundang Medical Center, CHA University, Seougnam, Korea. shin029@chamc.co.kr
  • 2Department of Urology, CHA Bundang Medical Center, CHA University, Seougnam, Korea.

Abstract

PURPOSE
The present study compared the difference between intraoperative transrectal ultrasound (iTRUS)-based prostate volume and preplan computed tomography (CT), preplan magnetic resonance imaging (MRI)-based prostate volume to estimate the number of seeds needed for appropriate dose coverage in permanent brachytherapy for prostate cancer.
MATERIALS AND METHODS
Between March 2007 and March 2011, among 112 patients who underwent permanent brachytherapy with 125I, 60 image scans of 56 patients who underwent preplan CT (pCT) or preplan MRI (pMRI) within 2 months before brachytherapy were retrospectively reviewed. Twenty-four cases among 30 cases with pCT and 26 cases among 30 cases with pMRI received neoadjuvant hormone therapy (NHT). In 34 cases, NHT started after acquisition of preplan image. The median duration of NHT after preplan image acquisition was 17 and 21 days for cases with pCT and pMRI, respectively. The prostate volume calculated by different modalities was compared. And retrospective planning with iTRUS image was performed to estimate the number of 125I seed required to obtain recommended dose distribution according to prostate volume.
RESULTS
The mean difference in prostate volume was 9.05 mL between the pCT and iTRUS and 6.84 mL between the pMRI and iTRUS. The prostate volume was roughly overestimated by 1.36 times with pCT and by 1.33 times with pMRI. For 34 cases which received NHT after image acquisition, the prostate volume was roughly overestimated by 1.45 times with pCT and by 1.37 times with pMRI. A statistically significant difference was found between preplan image-based volume and iTRUS-based volume (p < 0.001). The median number of wasted seeds is approximately 13, when the pCT or pMRI volume was accepted without modification to assess the required number of seeds for brachytherapy.
CONCLUSION
pCT-based volume and pMRI-based volume tended to overestimate prostate volume in comparison to iTRUS-based volume. To reduce wasted seeds and cost of the brachytherapy, we should take the volume discrepancy into account when we estimate the number of 125I seeds for permanent brachytherapy.

Keyword

Brachytherapy; Prostate cancer; Prostate volume

MeSH Terms

Brachytherapy
Humans
Magnetic Resonance Imaging
Prostate
Prostatic Neoplasms
Retrospective Studies
Seeds
Full Text Links
  • ROJ
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr