Korean J Med Phys.  2009 Dec;20(4):290-297.

Study on the Small Fields Dosimetry for High Energy Photon-based Radiation Therapy

Affiliations
  • 1Radiation Interactions and Dosimetry Lab, Department of Nuclear Engineering, Hanyang University, Seoul, Korea.
  • 2Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. youngyih@skku.edu
  • 3School of Electrical Engineering and Computer Science, Kyungpook National University, Daegu, Korea.

Abstract

In case of radiation treatment using small field high-energy photon beams, an accurate dosimetry is a challenging task because of dosimetrically unfavorable phenomena such as dramatic changes of the dose at the field boundaries, dis-equilibrium of the electrons, and non-uniformity between the detector and the phantom materials. In this study, the absorbed dose in the phantom was measured by using an ion chamber and a diode detector widely used in clinics. GAFCHROMIC(R) EBT films composed of water equivalent materials was also evaluated as a small field detector and compared with ionchamber and diode detectors. The output factors at 10 cm depth of a solid phantom located 100 cm from the 6 MV linear accelerator (Varian, 6 EX) source were measured for 6 field sizes (5x5 cm2, 2x2 cm2, 1.5x1.5 cm2, 1x1 cm2, 0.7x0.7 cm2 and 0.5x0.5 cm2). As a result, from 5x5 cm2 to 1.5x1.5 cm2 field sizes, absorbed doses from three detectors were accurately identified within 1%. Wheres, the ion chamber underestimated dose compared to other detectors in the field sizes less than 1x1 cm2. In order to correct the observed underestimation, a convolution method was employed to eliminate the volume averaging effect of an ion chamber. Finally, in 1x1 cm2 field the absorbed dose with a diode detector was about 3% higher than that with the EBT film while the dose with the ion chamber after volume correction was 1% lower. For 0.5x0.5 cm2 field, the dose with the diode detector was 1% larger than that with the EBT film while dose with volume corrected ionization chamber was 7% lower. In conclusion, the possiblity of GAFCHROMIC(R) EBT film as an small field dosimeter was tested and further investigation will be proceed using Monte Calro simulation.

Keyword

Small field; Absorbed dose; GAFCHROMIC(R) EBT; Ion chamber; Diode detector

MeSH Terms

Electrons
Particle Accelerators
Water
Water
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr