Prog Med Phys.  2013 Sep;24(3):140-144. 10.14316/pmp.2013.24.3.140.

Reference Dosimetry and Calibration of Glass Dosimeters for Cs-137 Gamma-rays

Affiliations
  • 1Research Center, Dongnam Institute of Radiological & Medical Sciences, Busan, Korea. physics7@empal.com
  • 2Department of Materials Physics, Dong-A University, Busan, Korea.

Abstract

In this research, the glass dosimeter was calibrated to measure the standard absorbed dose of the Cs-137 irradiator and absorbed dose in a biological sample. Absorbed dose in water for Cs-137 gamma ray was determined by the IAEA TRS-277 protocol. The PTW-TM30013 ion chamber and the PTW-TM41023 water phantom were utilized for measuring absorbed dose and the value was compared with the reading from DoseAce GD-302M glass dosimeter from Asahi Techno Glass Corporation for its calibration. The uncertainty of measurement (1sigma) of the calibrated glass dosimeter was 2.7% and this result would be applied to improve the accuracy in measurement of absorbed dose in a biological sample.

Keyword

Cs-137 irradiator; TRS-277 protocol; Glass dosimeter

MeSH Terms

Calibration*
Gamma Rays
Glass*
Uncertainty
Water
Water

Figure

  • Fig. 1. A photography of geometrical setup for measuring absorbed dose to water for a Cs-137 gamma-ray irradiator.

  • Fig. 2. A glass dosimeter located at a depth in water using Markus chamber adaptor.


Reference

1. Jeong DH, Kim JK, Yang KM, et al. Absorbed dose determination for a biological sample irradiated by gamma rays from a Cs-137 source. Korean J Med Phys. 22(3):124–130. 2011.
2. Sabine H. Methods in Radiobiology. Nanodosimetry Workshop. 2009.
3. Kwak DH, Moon YM, Kang YR, Lee CY, Jeong DH, Kim JK. 감마선 조사시 용기의 크기에 따른 흡수선량 평가. KARP 춘계학술대회 논문집. 2011, Seoul, pp.282.
4. Brady SL, Toncheva G, Dewhirst MW, Yoshizumi TT. Characterization of a 137Cs irradiator from a new perspective with modern dosimetric tools. Health Phys. 97(3):195–205. 2009.
Article
5. TRS-277. Absorbed Dose Determination in Photon and Electron Beams: An International Code of Practice. International Atomic Energy Agency, Vienna. 1997.
6. TRS-398. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. International Atomic Energy Agency, Vienna. 2000.
Full Text Links