Korean J Med Phys.  2008 Sep;19(3):164-171.

Analysis of Low MU Characteristics of Siemens Primus Linear Accelerator using Diode Arrays for IMRT QA

Affiliations
  • 1Department of Radiation Oncology, School of Medicine, Ewha Womans University, Seoul, Korea. renalee@ewha.ac.kr

Abstract

One of the most important task in commissioning intensity modulated radiotherapy (IMRT) into a clinic is the characterization of dosimetry performance under small monitor unit delivery conditions. In this study, method of evaluating dose monitor linearity, beam flatness and symmetry, and MLC positioning accuracy using a diode array is investigated. Siemens Primus linear accelerator (LA) with 6 and 10 MV x-rays was used to deliver radiation and the characteristics were measured using a multi array diodes. Monitor unit stabilities were measured for both x-ray energies. The dose linearity errors for the 6 MV x-ray were 2.1, 3.4, 6.9, 8.6, and 15.4 % when 20 MU, 10 MU, 5 MU, 4 MU, and 2 MU was delivered, respectively. Greater errors were observed for 10 MV x-rays with a maximum of 22% when 2 MU was delivered. These errors were corrected by adjusting D1_C0 values and reduced to less than 2% in all cases. The beam flatness and symmetry were appropriate without any correction. The picket fence test performed using diode array and film measurement showed similar results. The use of diode array is a convenient method in characterizing beam stability, symmetry and flatness, and positioning accuracy of MLC for IMRT commissioning. In addition, adjustment of D1-C0 value must be performed when a Siemens LA is used for IMRT because factory value usually gives unacceptable beam stability error when the MU/segment is smaller than 20.

Keyword

IMRT; Small MU; Symmetry; Flatness

MeSH Terms

Organothiophosphorus Compounds
Particle Accelerators
Organothiophosphorus Compounds
Full Text Links
  • KJMP
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr