Anesth Pain Med.  2023 Jan;18(1):46-50. 10.17085/apm.22212.

Secondary migration of a pre-existing central venous catheter due to a Swan-Ganz catheter insertion – A case report –

Affiliations
  • 1Department of Anesthesiology and Pain Medicine, Inje University Sanggye Paik Hospital, Seoul, Korea

Abstract

Background
The entanglement of multiple central venous catheters is a rare and seriouscomplication. The Swan-Ganz catheter is a responsible for various cases.Case: A 66-year-old male patient was under general anesthesia for a coronary artery bypassgraft surgery. As he had a pre-existing Perm catheter in the right subclavian vein, a SwanGanz catheter was inserted into the left internal jugular vein. Chest radiograph after catheterplacement revealed that the Perm catheter had migrated to the left brachiocephalic vein.The surgeon attempted to reposition it manually, but postoperative radiograph showed thatit had rolled into a loop. On postoperative day 1, radiological intervention was performed tountangle the loop, which was successful.Conclusions: After placing a Swan-Ganz catheter in patients with a pre-existing central venous catheter, the presence of entanglement should be assessed. In such cases, radiology-guided correction is recommended, as a blind attempt to disentangle can aggravate thecondition.

Keyword

Anesthesia, cardiac procedures; Catheters, indwelling; Catheterization, Swan-Ganz; Central venous catheters; Equipment failure; Intraoperative complications
Full Text Links
  • APM
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2023 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr