Lab Med Online.  2020 Jan;10(1):39-45. 10.3343/lmo.2020.10.1.39.

Comparison of Serum Creatinine Measurements among Roche Modular D, Cobas 8000 c702, and Beckman Coulter AU5800, by Jaffe and Enzymatic Methods

Affiliations
  • 1Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. nayadoo@hanmail.net

Abstract

BACKGROUND
Creatinine (Cr) is a representative biomarker reflecting renal function. In this study, we compared serum Cr levels using Roche Modular D (Roche Diagnostics, Germany), Roche Cobas 8000 c702 (Roche Diagnostics), and AU5800 (Beckman Coulter, USA). In addition, we assessed the differences in Cr measurements using the Jaffe and enzymatic methods.
METHODS
Precision, linearity, and methods were evaluated in accordance with CLSI guidelines. Serum Cr was measured by Modular D following the Jaffe method, and serum Cr was measured by Cobas 8000 c702 and AU5800, following the Jaffe and enzyme methods.
RESULTS
All of the total coefficients of variations (CVs) were below 5%. Linearity was observed in the performance ranges evaluated (r>0.99, slope: 0.965 and 0.955). When Modular D and Cobas 8000c 702 were compared, the slope and y-intercept were 0.9928 (95% confidence interval [CI]: 0.9802 to 1.000) and -0.0156 (95% CI: −0.0200 to −0.0054), respectively. The slope and y-intercept were 0.9811 (95% CI: 0.9570 to 0.9951) and -0.0484 (95% CI: −0.0638 to −0.0297) when Modular D and Au5800 were compared. Serum Cr measured by Cobas 8000 c702 and AU5800 using the Jaffe method were 3.2% and 6.9% lower than the values measured by Modular D, respectively. Both Modular D and Cobas 8000 c702 showed acceptable accuracies.
CONCLUSIONS
Serum Cr measurements using Cobas 8000 c702 and AU5800 were comparable to those measured by Modular D, and showed satisfactory precision and linearity; thus, these techniques could be useful for clinical laboratories.

Keyword

Creatinine; Enzyme method; Jaffe method; Modular D; Cobas 8000 c702; AU5800

MeSH Terms

Creatinine*
Methods*
Creatinine

Figure

  • Fig. 1 Scatter plots and difference plots comparing creatinine measurements obtained using Modular D (Jaffe) with those obtained using Cobas 8000 c702 (Jaffe and enzymatic) and AU5800 (Jaffe and enzymatic). (A) Modular D with Cobas 8000 c702 (Jaffe), (B) Modular D with AU5800 (Jaffe), (C) Modular D with Cobas 8000 c702 (enzymatic), (D) Modular D with Au5800 (enzymatic), (E) Cobas 8000 c702 (Jaffe) with Cobas 8000 c702 (enzymatic), (F) AU5800 (Jaffe) with AU5800 (enzymatic).


Reference

1. Stevens LA, Coresh J, Greene T, Levey AS. Assessing kidney function–measured and estimated glomerular filtration rate. N Engl J Med. 2006; 354:2473–2483.
Article
2. Stevens LA, Levey AS. Measured GFR as a confirmatory test for estimated GFR. J Am Soc Nephrol. 2009; 20:2305–2313.
Article
3. Welch MJ, Cohen A, Hertz HS, Ng KJ, Schaffer R, Van der, et al. Determination of serum creatinine by isotope dilution mass spectrometry as a candidate definitive method. Anal Chem. 1986; 58:1681–1685.
Article
4. Armbruster D, Miller RR. The Joint Committee for Traceability in Laboratory Medicine (JCTLM): a global approach to promote the standardisation of clinical laboratory test results. Clin Biochem Rev. 2007; 28:105–113.
5. Peake M, Whiting M. Measurement of serum creatinine–current status and future goals. Clin Biochem Rev. 2006; 27:173–184.
6. Miller WG, Myers GL, Ashwood ER, Killeen AA, Wang E, Thienpont LM, et al. Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. Arch Pathol Lab Med. 2005; 129:297–304.
Article
7. Myers GL, Miller WG, Coresh J, Fleming J, Greenberg N, Greene T, et al. Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. Clin Chem. 2006; 52:5–18.
Article
8. Clinical and Laboratory Standards Institute. Evaluation of precision of quantitative measurement procedures; Approved guideline-Third edition. CLSI document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute;2014.
9. Clinical and Laboratory Standards Institute. Evaluation of the linearity of quantitative measurement procedures: A statistical approach; Approved guideline. CLSI document EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute;2003.
10. Clinical and Laboratory Standards Institute. Measurement procedure comparison and bias estimation using patients samples; Approved guideline-Third editon. CLSI document EP09-A3. Wayne, PA: Clinical and Laboratory Standards Institute;2013.
11. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et al. Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation. https://www.westgard.com/biodatabase1.htm (Updated on 2014).
12. White GH, Farrance I. AACB Uncertainty of Measurement Working Group. Uncertainty of measurement in quantitative medical testing: a laboratory implementation guide. Clin Biochem Rev. 2004; 25:S1–24.
13. Biswas SS, Bindra M, Jain V, Gokhale P. Evaluation of imprecision, bias and total error of clinical chemistry analysers. Indian J Clin Biochem. 2015; 30:104–108.
Article
14. Delanghe JR, Cobbaert C, Harmoinen A, Jansen R, Laitinen P, Panteghini M. Focusing on the clinical impact of standardization of creatinine measurements: a report by the EFCC Working Group on Creatinine Standardization. Clin Chem Lab Med. 2011; 49:977–982.
Article
15. Killeen AA, Ashwood ER, Ventura CB, Styer P. Recent trends in performance and current state of creatinine assays. Arch Pathol Lab Med. 2013; 137:496–502.
Article
16. Statland BE. Clinical decision levels for lab tests. 2nd ed. Oradell, NJ: Medical Economics Books;1987. p. 72–73.
17. Jun SH, Song J, Song WH. Annual report on the external quality assessment scheme for clinical chemistry in Korea (2015). J Lab Med Qual Assur. 2016; 38:111–119.
Article
18. Weber JA. Interferences in current methods for measurements of creatinine. Clin Chem. 1991; 37:695–700.
Article
19. Greenberg N, Roberts WL, Bachmann LM, Wright EC, Dalton RN, Zakowski JJ, et al. Specificity characteristics of 7 commercial creatinine measurement procedures by enzymatic and Jaffe method principles. Clin Chem. 2012; 58:391–401.
Article
20. O'Leary N, Pembroke A, Duggan PF. A simplified procedure for eliminating the negative interference of bilirubin in the Jaffé reaction for creatinine. Clin Chem. 1992; 38:1749–1751.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr