Lab Med Online.  2019 Jan;9(1):6-11. 10.3343/lmo.2019.9.1.6.

Performance Evaluation of the CRE2 Reagent from Siemens for Serum Creatinine Measurement

Affiliations
  • 1Department of Laboratory Medicine, Chungnam National University Hospital, Daejeon, Korea. kckwon@cnu.ac.kr

Abstract

BACKGROUND
For creatinine measurement, the enzymatic method is known to be more accurate than the Jaffe method; however, the latter is still widely used. We evaluated the performance of the CRE2 reagent (Siemens Healthcare Diagnostics Inc., USA), which uses a modified Jaffe method.
METHODS
Three quality control standards were used for precision evaluations of CRE2 on Dimension VISTA 500 instrument (Siemens). Moreover, the linearity and carryover characteristics were assessed. Sixty-eight creatinine results obtained using the CRE2 and ECREA (enzymatic) reagents (Siemens) were compared with those obtained using the L-CRE (enzymatic) reagent (Shinyang Diagnostics, Korea). The accuracy of CRE2, ECREA, and L-CRE was evaluated using a standard reference material.
RESULTS
The CV of within-run (0.7-2.4%), between-run (0.4-1.7%), between-day precision (0.7-0.9%) for three standards, and total CV for medium (1.6%) and high levels (1.3%) satisfied the analytical goal. The linearity for CRE2 was excellent (R2=0.999). Comparisons of CRE2 and ECREA to L-CRE were well correlated (r=0.996 and 0.997, respectively). In comparison with L-CRE, 5 CRE2 results and 15 ECREA results exceeded minimum bias goal (5.1%) in samples with creatinine levels of >1 mg/dL. The carryover rate was −0.04%. In terms of accuracy, the percent bias values of CRE2, ECREA, and L-CRE were 7.4, −6.4, and −3.4, respectively, for low level; and 3.9, −1.5, and 0.7, respectively, for high level.
CONCLUSIONS
For creatinine measurements, the CRE2 reagent showed good performance. It can be used in the diagnosis, treatment monitoring, and risk assessment of kidney diseases.

Keyword

CRE2; Precision; Linearity; Comparison; Accuracy

MeSH Terms

Bias (Epidemiology)
Creatinine*
Delivery of Health Care
Diagnosis
Indicators and Reagents
Kidney Diseases
Methods
Quality Control
Risk Assessment
Creatinine
Indicators and Reagents

Figure

  • Fig. 1. Comparison of creatinine values measured by L-CRE and CRE2 reagents.

  • Fig. 2. Comparison of creatinine values measured by L-CRE and ECREA reagents.


Reference

1.Chung HJ., Chun SI., Min WK. Creatinine determination with minimized interference. J Lab Med Qual Assur. 2008. 30:229–31.
2.Burtis CA., Ashwood ER, et al. eds. Tietz textbook of clinical chemistry and molecular diagnostics. 4th ed.St. Louis, MO: Elsevier Inc.;2006. p. 797–801.
3.O'Leary N., Pembroke A., Duggan PF. A simplifed procedure for eliminating the negative interference of bilirubin in the Jaffe reaction for creatinine. Clin Chem. 1992. 38:1749–51.
4.Myers GL., Miller WG., Coresh J., Fleming J., Greenberg N., Greene T, et al. Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. Clin Chem. 2006. 52:5–18.
Article
5.Siemens Dimension Vista® system Creatinine (CRE2) Flex® reagent cartridge. https://www.accessdata.fda.gov/cdrh_docs/pdf13/K133728.pdf.
6.Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples; approved guideline. 2nd ed.EP09-A2. Wayne, PA: Clinical and Laboratory Standard Institute;2002.
7.Clinical and Laboratory Standards Institute. Evaluation of precision of quantitative measurement procedures; approved guideline. 3rd ed.EP05-A3. Wayne, PA: Clinical and Laboratory Standard Institute;2014.
8.Clinical and Laboratory Standards Institute. Evaluation of the linearity of quantitative measurement procedures; a statistical approach; approved guideline. EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute. 2003.
9.Clinical and Laboratory Standards Institute. Preliminary evaluation of quantitative clinical laboratory measurement procedures; approved guideline. 3rd ed.EP10-A3.Wayne, PA: Clinical and Laboratory Standards Institute;2006.
10.Robert L. National Institute of Standards and Technology certifcate of analysis, standard reference material. 665:2014. 1–3.
11.Nah H., Lee SG., Lee KS., Won JH., Kim HO., Kim JH. Evaluation of bilirubin interference and accuracy of six creatinine assays compared with isotope dilution-liquid chromatography mass spectrometry. Clin Bio-chem. 2016. 49:274–81.
Article
12.Ricos C., Alvarez V., Cava F., Garcia-Lario JV., Hernandez A., Jimenez CV, et al. Desirable specifcations for total error, imprecision, and bias, derived from intra- and inter-individual biological variation. https://www.westgard.com/biodatabase1.htm. (updated on 2014).
13.Burtis CA., Ashwood ER, et al. eds. Tietz textbook of clinical chemistry and molecular diagnostics. 5th ed.St. Louis, Mo: Elsevier Inc;2006. p. 196–8.
14.Hermida FJ., Lorenzo MJ., Pérez A., Fernández M., Sagastagoia O., Maga-dán C. Comparison between ADVIA Chemistry systems Enzymatic Creatinine_2 method and ADVIA Chemistry systems Creatinine method (kinetic Jaffe method) for determining creatinine. Scand J Clin Lab Invest. 2014. 74:629–36.
Article
15.Schmidt RL., Straseski JA., Raphael KL., Adams AH., Lehman CM. A risk assessment of the Jaffe vs enzymatic method for creatinine measurement in an outpatient population. PLoS ONE. 2015. 10:e0143205.
Article
16.Panteghini M and IFCC Scientifc Division. Enzymatic assays for creatinine: time for action. Clin Chem Lab Med. 2008. 46:567–72.
Full Text Links
  • LMO
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr