J Lab Med Qual Assur.  2006 Dec;28(2):225-227.

Comparison of Sekisui Trank Insepack and BD Vacutainer Plastic Citrate Tubes for Routine Coagulation Assays

Affiliations
  • 1Department of Laboratory Medicine, Hallym University College of Medicine, Seoul, Korea. parkjy@hallym.or.kr

Abstract

BACKGROUND Coagulation assays can be easily affected by preanalytical variables, including the factors associated with the blood collection tube. Recently the Sekisui Trank Insepack (Sekisui Chemical Co., Ltd, Osaka, Japan; Insepack) blood collection tubes were introduced into Korea. The aim of this study was to compare the results of routine coagulation assays using the Insepack with those using BD Vacutainer (BD, Franklin Lakes, NJ, USA; Vacutainer) tubes, which are most widely used in Korea.
METHODS
Paired blood samples from 64 patients were drawn into Insepack and Vacutainer 3.2% citrate tubes. The results of prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, and fibrin/fibrinogen degradation product (FDP) were determined in parallel.
RESULTS
No statistically significant differences were observed between Insepack and Vacutainer tubes for PT, fibrinogen, and FDP tests. For APTT which showed the significant differences between the results in the two tubes the mean difference was small (0.8 seconds), and it was not considered clinically significant.
CONCLUSIONS
The results of this study suggest that Insepack tubes can be used in place of Vacutainer tubes for routine coagulation assays.

Keyword

Sekisui Trank Insepack; BD Vacutainer; Routine Coagulation Assay

MeSH Terms

Citric Acid*
Fibrinogen
Humans
Japan
Korea
Lakes
Partial Thromboplastin Time
Plastics*
Prothrombin Time
Citric Acid
Fibrinogen
Plastics
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr