J Lab Med Qual Assur.  2008 Jun;30(1):189-194.

Evaluation of CELL-DYN Sapphire Hematology Analyzer

Affiliations
  • 1Department of Laboratory Medicine, College of Medicine, Kyung Hee University, Seoul, Korea. peanut500@hanmail.net

Abstract

BACKGROUND: CELL-DYN Sapphire (Abbott Diagnostics, Santa Clara, CA, USA) hematology analyzer is based on laser multi angle polarized scatter separation, focused flow impedance and fluorescent flow cytometry. In order to assess the usefulness of this newly introduced analyzer, we evaluated the precision and correlation of complete blood count including white blood cell differential count and reticulocyte count of CELL-DYN Sapphire instrument.
METHODS
Patient samples ordered for complete blood count and white blood cell differential count and CELL-DYN 29 Plus control samples were used. We evaluated the precision of complete blood count and WBC differential count and analyzed correlation of these parameters for comparison with ADVIA 120 (Bayer corporation, Tarrytown, NY, USA) and manual count. We additionally evaluated reticulocyte parameters for precision and comparison with ADVIA 120.
RESULTS
Short term precisions showed low coefficient of variation (CV) values for all CBC parameters including lower than 3% CV values for neutrophil and lymphocyte count. Long term precision showed lower than 4% CV values for all CBC and WBC differential count except for basophils. Comparison with ADVIA 120 showed good correlation for CBC except for MCHC. Good correlation was confirmed in all differential counts except for basophils. Comparison with manual WBC differential count showed good correlation except for monocyte and eosinophil.
CONCLUSIONS
CELL-DYN Sapphire has high precision and good correlation with ADVIA 120 and manual count. CELL-DYN Sapphire is an exceptional instrument for CBC with WBC differentials and reticulocyte counting.

Keyword

CELL-DYN Sapphire; ADVIA 120; Manual count; Precision; Correlation

MeSH Terms

Aluminum Oxide
Basophils
Blood Cell Count
Electric Impedance
Flow Cytometry
Hematology
Humans
Leukocytes
Lymphocyte Count
Monocytes
Neutrophils
Reticulocyte Count
Reticulocytes
Aluminum Oxide
Full Text Links
  • JLMQA
Actions
Cited
CITED
export Copy
Close
Share
  • Twitter
  • Facebook
Similar articles
Copyright © 2024 by Korean Association of Medical Journal Editors. All rights reserved.     E-mail: koreamed@kamje.or.kr