J Korean Soc Ther Radiol.  1995 Mar;13(1):101-112.

Calibration and Radiation Survey of High Dose Rate Remote Afterloading System

Affiliations
  • 1Department of Radiation Oncology, Wonkwang University, School of Medicine, Korea.
  • 2Department of Radiation Oncology, Chonju Presbyterian Medical Center, Korea.

Abstract

High Dose Rate Remote Afterloading system was installed at Wonkwang Universi-ty Hospital in January 1994. In this report, the calibration of a Gammamed 12-i High Dose Rate Remote Afterloading system and the radiation survey around the facility after design and construct a shieding room are discussed. The radiation survey of the facility indicates that the use of ordinary concrete shielding of existing room will provide adequate shielding. Also, the methodologies for performing source calibra-tion are presented.

Keyword

Aquaplast; Surface dose; 6MV X-ray; Skin-sparing effect